LumiCam 1300 Advanced - for extremely high requirements

Instrument Systems developed the LumiCam 1300 Advanced with the objective of further enhancing the measuring accuracy of the LumiCam 1300 Color, especially regarding color light and narrowband LED color light.

This instrument is based on an innovative concept with 6 filters. Alongside the four optical filters of the LumiCam 1300 Color (X'1, X'2, Y', Z'), two additional filters (K', L') are used. In combination with an optimized algorithm, the tristimulus values are calculated much more accurately compared with previous methods.

Measurement of the color coordinates is always critical and has been significantly improved with this patent-pending procedure. The LumiCam 1300 Advanced is therefore the ideal measuring instrument for all applications where precise determination of color coordinates is a top priority.

The LumiCam 1300 Advanced is compatible with the complete range of lenses and accessories in the LumiCam 1300 Family.

All functions of the LumiCam software are also available for the LumiCam 1300 Advanced. If required, it can also be operated like a LumiCam 1300 Mono or Color – with analog measuring times and accuracies. The relevant operating mode can be easily selected in the software.

Technical Specifications:

 

LumiCam 1300 Color

LumiCam 1300 Advanced

Effective pixels

1370 (H) x 1020 (V)

Measurement range

0.0001 to 100,000 cd/m²
higher values with optional density filters
(density 2 and density 4)

Photometric measured parameters

Luminance, luminous intensity, contrast

Colorimetric measured parameters

Color coordinates x, y, and u',v'
Dominant wavelength
Color saturation
Correlated color temperature

Linearity

+/- 0.5%

Luminance accuracy for standard illuminant A

+/- 4%

Luminance accuracy for LED color light

n.a.

+/- 2%

Accuracy color coordinates for std. illuminant A

+/-0.003

+/-0.003

Accuracy color coordinates for color light

+/-0.010

+/-0.003

Accuracy color coordinates for LED color light

n.a.

+/-0.005

Dimensions

275 mm x 120 mm x 175 mm

Weight

3 kg

List of Lenses:

Order Number

LC-312

LC-312

LC-314

LC-316-1

Focal length

14 mm

28 mm

50 mm

105 mm

Minimum focusing distance
(DUT - lens)

11 cm

18 cm

37 cm

11 cm

Minimum measurement distance
(DUT - housing)

25 cm

25 cm

50 cm

30 cm

Image size  at min. measurement distance (h x v)

116 x 86 mm

64 x 48 mm

77 x 57 mm

14 x 10 mm

Pixel size at min. measurement distance

85 µm

47 µm

56 µm

10 µm

Image size at 1 m distance (h x v)

570 x 425 mm

300 x 220 mm

160 x 120 mm

80 x 60 mm

Pixel size at 1 m distance

416 µm

217 µm

119 µm

58 µm

6-filter technology

The LumiCam 1300 Advanced records the data in six channels (X'1, X'2, Y', Z', K', L') to determine the three tristimulus values (X, Y, Z). This results in a redundancy for their calculation. Using six filters effectively makes adjustment to the sensitivity curves more accurate and is implemented in practical terms by calculating the tristimulus values from multiplication of a 3x6 adaptation matrix with the six output channels. The adaptation matrix is determined individually for each LumiCam 1300 Advanced in the calibration process and optimizes the accuracy of the relevant quantities.