DTS 140 - a universal system for display measurement and more
Complete systems based on the CAS 140CT array spectrometer featuring a high-end, back-illuminated CCD detector are available for measurements of radiance and luminance as well as color of displays and back-illuminated panel graphics. No matter what kind of display technology has to be tested - whether LCD, CRT, OLED or LED displays - the DTS 140 can be used universally.
Coupling of light into the system is achieved using the TOP 200 telescopic optical probe that is attached to the spectrometer via an optical fiber. A large selection of lenses is available for the TOP 200 in order to provide different measuring spot sizes. The new SpecWin Pro software offers comprehensive tools for data analysis and documentation of measurement results.
The fiber coupling of the TOP 200 makes it possible to simply swap the adapter and quickly reconfigure the DTS140 for other applications (e.g. LED measurement). The range of applications available is thus considerably expanded - a feature hard to realize for spectrometers with a fixed, integrated lens.
Features:
very accurate measurements combined with high sensitivity
modified Pritchard optics guarantees ideal positioning of the measurement spot
integrated alignment camera with wide field of view
apertures and optical density filters selectable via software
new SpecWin Pro software provides for convenient analysis and documentation of results
Specifications
DTS140-131
DTS140-133
DTS140-231
DTS140-233
Spectral range
360 - 830 nm
380 - 1040 nm
360 - 830 nm
380 - 1040 nm
Spectral resolution
2.2 nm
3.0 nm
2.2 nm
3.0 nm
Datapoint interval
0.5 nm
0.65 nm
0.5 nm
0.65 nm
TOP 200 Telescopic Optical Probe
Lens
60 mm
60 mm
HRL 90
HRL 90
Working distance *)
18 cm
18 cm
23 cm
23 cm
Stray light (approx.)
1 %
1 %
0.1 %
0.1 %
Measurement sensitivity range [cd/m²]
Aperture 1
0.2 - 1*109
0.1 - 1*109
0.6 - 4*109
0.4 - 3*109
Aperture 2
0.06 - 4*108
0.04 - 3*108
0.2 - 1*109
0.1 - 3*109
Aperture 3
0.02 - 1*108
0.01 - 1*108
0.06 - 4*108
0.04 - 3*108
Aperture 4
0.006 - 4*107
0.004 - 3*107
0.02 - 1*108
0.01 - 1*108
Accuracy
Luminance
± 3 %
Radiance
± 4 %
Color coordinates (x,y)
± 0.0015
Dominant wavelength
± 0.5 nm
Polarization sensitivity
< 1 %
*) Distance from the test sample to the TOP 200's front panel; for models with the 60 mm lens, basic calibration is done at a working distance of 50 cm.